WebSelect the CytoFLEX Ready to Use (RTU) Daily QC Fluorospheres Target Values for correct lot number of Ready to Use (RTU) QC Fluorospheres and then select Download and save to desired file path or USB drive. 7. If QC fails, select Daily Clean and Prime from the Cytometer menu, follow the prompts and repeat QC. If QC fails two WebThe CytoFLEX system brings you easily upgradeable detection capabilities for up to 3 lasers and 13 color research flow cytometry right on your bench top. Pro...
CytoFlex Quick Guide - Beth Israel Deaconess Medical Center
WebOct 7, 2024 · On the Daily Clean window, select the desired wells that contain the cleaning solution (three wells) and the wells that contain water (five wells). Note: User Coulter Clenz as cleaning solution and PureLab diH2O as water. 2 Place the plate back on the plate loader and select Load and then Start to start the cleaning protocol. 3 WebGo to Cytometer > System Start up Program The plate loader automatically ejects the plate holder stage and the System Startup Program window appears. Follow the on screen … how to save a sound
CytoFlex User Guide - NYU Langone Health
WebCytoFLEX Maintenance-Every Day: o Daily Clean (using Flow-Clean and water): After QC, in between experiments, and at the end of the day -Once a month: o Deep clean = cleans flow cell = uses 50% Contrad 70 solution Shouldn’t leave Deep Clean solution in the flow cell for more than 24h WebPractitioner Application. Practitioner Application; Research & Instructions; Articles; Contact. Consumers: Contact Us; Shop. Products. CytoDetox® CytoDetox+® With … WebFeb 23, 2024 · To resolve this problem, we perform a daily clean using at least 6 wells of the cleanse solution, followed by 12 wells of water after each QC run. Each well is run for 1 min at a minimum of 160 uL flow rate. This daily clean can also be run as an experiment so that the level of bead contamination can be monitored. This process has worked well ... north face aksu